RULA technologies have created a specialized calibration software module, which can be used together with RL-C21 vibration control system.

What is an accelerometer

An accelerometer is a complicated electro-mechanical system, which usually contains a spring-mass system (piezoceramics and a seismic mass) and a build-in charge converter. It may also contain a memory chip (if it is a TEDS-sensor).
As every complicated system, an accelerometer needs to be periodically checked (usually one time per year) to ensure the following:
  • The sensor performs according to its specifications;
  • The sensor is still «healthy».
The calibration procedure is a requirement, however, it is not as simple as it seems.

What is an accelerometer

Calibration standards

Calibration is a strictly standardized procedure. The main standards referring to calibration procedures are the: 
  • ISO 16063-1:1998 Methods for the calibration of vibration and shock transducers — Part 1: Basic concepts
  • ISO 16063-12:2002 Methods for the calibration of vibration and shock transducers — Part 12: Primary vibration calibration by the reciprocity method
  • ISO 16063-21:2003 Methods for the calibration of vibration and shock transducers — Part 21: Vibration calibration by comparison to a reference transducer
  • ISO 16063-31:2009 Methods for the calibration of vibration and shock transducers — Part 31: Testing of transverse vibration sensitivity.

Calibration standards

Calibration Equipment

Calibration requires a set of equipment: 
  • a reference transducer;
  • a shaker with a power amplifier;
  • a specialized controller with precise input and output channels;
  • a PC with calibration software.

The typical calibration set is shown in the picture. It includes a reference sensor mounted on the shaker, a shaker and an amplifier.
If the sensor under calibration is mounted directly on the reference sensor, such a procedure is called back-to-back calibration.

Calibration Equipment

Calibration Procedure

The powerful software option, which allows calibrating accelerometers, can be run using RL-C21 and RL-C25 devices.
The software runs the procedure of automated calibration by comparison to a reference transducer.
The procedure contains two steps:
  • Reference sensitivity measurements. During the operation, we determine the sensitivity on the reference frequency (usually 160 or 200 Hz);
  • Frequency response measurements. During the operation, we determine the  sensitivity on all the frequencies of the sensors frequency range.

Calibration Procedure

Measurements

Usually, a sine wave is used to excite both sensors.

When the sine wave is being generated on the shaker, the output signals of the reference sensor and sensor under calibration are measured.

After that, sensitivity is calculated by the formula:

Sc – sensitivity of the sensor under calibration;
Sr – sensitivity of the reference sensor;
Xc – output of the sensor under calibration;
Xr – output of the reference sensor.

The procedure repeats itself on five or six frequencies distributed through the frequency range of the sensor under calibration, one frequency is the minimum, one the maximum and one the reference frequency of the accelerometer.

Measurements

Faster calibration

In spite of the fact that with RULA equipment all the excitations and calculations are carried out automatically, without any input from the user, sine excitation procedure takes long time.

Therefore, in addition to sine wave excitation, TestUp software provides calibration with Random signal excitation. In this case, the shaker is excited with a random signal, which contains all the frequencies in the sensor frequency range. If a random signal is used, sensitivities on all the frequencies are measured simultaneously. This way reduces the calibration time to three or four minutes.

Faster calibration

Better accuracy

Traditional back-to back calibration can be inaccurate if the sensor under calibration is bigger or heavier than the reference sensor. The error grows with the increase of frequency. To avoid such an error, in TestUp software we implemented the procedure of calibration by substitution.
Besides correcting the errors caused by the sensor mass and sizes, calibration by substitution eliminates the errors caused by the non-linearity of measurement channels and reduces the uncertainty of measurements.

Better accuracy

Creating reports

After any calibration, TestUp automatically creates  a comprehensive report, which includes:

  • measurement results in the form of tables and graphs;
  • detailed information about the reference sensor;
  • calibration date and place, etc.

Creating reports

Research and development

Besides the automated measurements, a manual measurement mode is implemented in TestUp.
The mode allows specifying the frequency and amplitude (in acceleration, velocity or displacement units) manually and checking the response from the sensors. The mode is highly useful for research purposes.

Research and development

Innovative, yet easy-to-use software

RULA Technologies have developed an extensive subsystem of vibration transducer calibration.
It is an innovative, yet easy-to-use software, which provides automated calibration of a wide range of vibration sensors. The system uses a range of calibration methods, innovative and well-established,  according to ISO 16063-21.

Our advantages  are:

  • Compliance with international standards;
  • Innovative calibration methods;
  • Short calibration time;
  • Easy intuitive interface;
  • Comprehensive reports.

Innovative, yet easy-to-use software
Conclusion
Vibration sensor calibration module by RULA Technologies is an efficient and flexible tool, which make your calibration routine considerably faster and easier.